Implementation of remote video auditing with feedback and compliance for manual-cleaning protocols of endoscopic retrograde cholangiopancreatography endoscopes

    loading  Checking for direct PDF access through Ovid

Abstract

A pilot initiative to assess the use of remote video auditing in monitoring compliance with manual-cleaning protocols for endoscopic retrograde cholangiopancreatography (ERCP) endoscopes was performed. Compliance with manual-cleaning steps following the initiation of feedback was measured. A video feed of the ERCP reprocessing room was provided to remote auditors who scored items of an ERCP endoscope manual-cleaning checklist. Compliance feedback was provided in the form of reports and reeducation. Outcomes were reported as checklist compliance. The use of remote video auditing to document manual processing is a feasible approach and feedback and reeducation increased manual-cleaning compliance from 53.1% (95% confidence interval, 34.7-71.6) to 98.9% (95.0% confidence interval, 98.1-99.6).