Correlation of Impedance Threshold Device use during cardiopulmonary resuscitation with post-cardiac arrest Acute Kidney Injury

    loading  Checking for direct PDF access through Ovid

Abstract

Purpose

To assess whether use of Impedance Threshold Device (ITD) during cardiopulmonary resuscitation (CPR) reduces the degree of post-cardiac arrest Acute Kidney Injury (AKI), as a result of improved hemodynamics, in a porcine model of ventricular fibrillation (VF) cardiac arrest.

Methods

After 8 min of untreated cardiac arrest, the animals were resuscitated either with active compression-decompression (ACD) CPR plus a sham ITD (control group, n = 8) or with ACD-CPR plus an active ITD (ITD group, n = 8). Adrenaline was administered every 4 min and electrical defibrillation was attempted every 2 min until return of spontaneous circulation (ROSC) or asystole. After ROSC the animals were monitored for 6 h under general anesthesia and then returned to their cages for a 48 h observation, before euthanasia. Two novel biomarkers, Neutrophil Gelatinase-Associated Lipocalin (NGAL) in plasma and Interleukin-18 (IL-18) in urine, were measured at 2 h, 4 h, 6 h, 24 h and 48 h post-ROSC, in order to assess the degree of AKI.

Results

ROSC was observed in 7 (87.5%) animals treated with the sham valve and 8 (100%) animals treated with the active valve (P = NS). However, more than twice as many animals survived at 48 h in the ITD group (n = 8, 100%) compared to the control group (n = 3, 37.5%). Urine IL-18 and plasma NGAL levels were augmented post-ROSC in both groups, but they were significantly higher in the control group compared with the ITD group, at all measured time points.

Conclusion

Use of ITD during ACD-CPR improved hemodynamic parameters, increased 48 h survival and decreased the degree of post-cardiac arrest AKI in the resuscitated animals.