Validation of the Andon KD-5851 upper arm blood pressure monitor, for self-measurement according to the European Society of Hypertension International Protocol revision 2010


    loading  Checking for direct PDF access through Ovid

Abstract

ObjectiveThis study aimed to validate the Andon KD-5851 upper arm blood pressure (BP) monitor according to the European Society of Hypertension International Protocol (ESH-IP) revision 2010.MethodsA total of 33 eligible participants were included in the study. Sequential measurements of BPs were performed using a mercury sphygmomanometer and the device, and the data analysis was carried out following precisely the ESH-IP revision 2010.ResultsThe device had 82, 98, and 99 measurements within 5, 10, and 15 mmHg for systolic blood pressure and 85, 95, and 99 measurements for diastolic blood pressure, respectively. The average device–observer difference was −0.53±4.00 mmHg for systolic blood pressure and −1.15±4.06 mmHg for diastolic blood pressure. The device passed all the criteria according to the ESH-IP revision 2010.ConclusionAccording to the validation results on the basis of the ESH-IP revision 2010, the Andon KD-5851 upper arm BP monitor can be recommended for self/home measurement in adults.