Spectral Entropy and Bispectral Index as Measures of the Electroencephalographic Effects of Propofol

    loading  Checking for direct PDF access through Ovid

Abstract

Recently, Datex-Ohmeda introduced the Entropy Module™ for measuring depth of anesthesia. Based on the Shannon entropy of the electroencephalogram, state entropy (SE) and response entropy (RE) are computed. We investigated the dose-response relationship of SE and RE during propofol anesthesia in comparison with the Bispectral Index™ (BIS). Twenty patients were studied without surgical stimulus. Anesthesia was induced by a constant propofol infusion of 2000 mg/h (451 ± 77 μg·min−1·kg−1) via a large forearm vein. Propofol was infused until substantial burst suppression occurred (more than 50%) or mean arterial blood pressure decreased to <60 mm Hg. Hereafter, infusions were stopped until recovery of BIS values up to 60 was reached. Subsequently, the constant propofol infusion of 2000 mg/h was restarted to increase depth of anesthesia and again decreased (infusion was stopped) within the BIS value range of 40–60. The coefficient of determination (R2) and the prediction probability (PK) were calculated to evaluate the performance of SE, RE, and BIS to predict changing propofol effect-site concentrations. R2 values for SE, RE, and BIS of 0.88 ± 0.08, 0.89 ± 0.07, and 0.92 ± 0.06, respectively, were similar. The calculated PK values, however, revealed a significant difference between SE and RE compared with BIS, with PK = 0.77 ± 0.09, 0.76 ± 0.10, and 0.84 ± 0.06, respectively. BIS seems to show slight advantages in predicting propofol effect-site concentrations compared with SE and RE, as measured by PK but not as measured by R2.