Validation of the Andon KD5031 for clinical use and self-measurement according to the European Society of Hypertension International Protocol

    loading  Checking for direct PDF access through Ovid

Abstract

Objective

This study aimed to evaluate the accuracy of the automated oscillometric upper arm blood pressure (BP) monitor Andon KD5031 for home BP monitoring according to the European Society of Hypertension International Protocol revision 2010.

Methods

Systolic blood pressure (SBP) and diastolic blood pressure (DBP) were sequentially measured in 33 participants using the standard mercury sphygmomanometer and the Andon KD5031 device. Ninety-nine pairs of comparisons were obtained from 33 participants for analysis.

Results

The KD5031 device achieved the targets in part 1 of the validation study. The number of absolute differences between the device and the observers within a range of 5, 10, and 15 mmHg was 66/99, 93/99, and 98/99, respectively, for SBP and 72/99, 94/99, and 99/99, respectively, for DBP. The device also achieved the targets in part 2 of the validation study. Twenty-six participants for both SBP and DBP had at least two of the three device–observer differences within 5 mmHg (required ≥24). The number of participants without a device–observer difference within 5 mmHg was one for SBP and three for DBP (required ≤3).

Conclusion

The Andon upper arm BP monitor KD5031 has passed the International Protocol requirements, and it can be recommended for clinical use and self-measurement in adults.