Early malfunction of polyvinyl alcohol membrane of septal occluder

    loading  Checking for direct PDF access through Ovid

Abstract

Percutaneous device closure of atrial septal defect (ASD) is now considered the choice of treatment. Numerous devices with advantages/disadvantages are currently available and under development. Cardia Ultrasept II ASD occluder has a nitinol frame covered with polyvinyl alcohol (PVA) membrane. Here, a well-documented case of early malfunction of PVA membrane detected in the first week of implantation in a 4-year-old male patient, who underwent an uneventful device closure with 20 mm Cardia Ultrasept II ASD Occluder, is presented. One week after implantation left-to-right shunt through the device was detected and the explantation of device revealed PVA membrane with multiple perforations. © 2015 Wiley Periodicals, Inc.

Related Topics