Optical Coherence Tomography and Ultrasound Biomicroscopy Imaging of Opaque Corneas

    loading  Checking for direct PDF access through Ovid

Abstract

Purpose:

The purpose was to determine the interchangeability of ultrasound biomicroscopy (UBM) and anterior segment optical coherence tomography (AS-OCT) for corneal opacity depth measurement.

Methods:

Twenty-six eyes of 26 consecutive patients with corneal opacities were examined by both AS-OCT and UBM. The corneal thickness and the corneal opacity depth were measured and compared. The interchangeability was determined by Bland–Altman plotting.

Results:

The difference in the full corneal thickness and in the corneal opacity depth between OCT and UBM was 5 ± 7 μm and −1 ± 8 μm, respectively. There were strong correlations and no significant differences between the paired parameters (all r > 0.99, P < 0.01). The limits of agreement were 5 ± 14 μm for the corneal thickness and −1 ± 14.8 μm for the corneal opacity depth.

Conclusions:

AS-OCT and UBM may be used interchangeably for measuring both full corneal thickness and corneal opacity depth in patients with corneal opacity.