Trends in the Use of Pelvimetry Techniques

    loading  Checking for direct PDF access through Ovid

Abstract

Assessment of maternal pelvic dimensions is usually considered necessary where vaginal delivery is contemplated in a breech presentation or if reduced pelvic dimensions are suspected in a current or previous pregnancy. Pelvimetry techniques include computed tomography (CT), conventional radiography, digital fluorography and magnetic resonance imaging (MRI). The first three techniques result in a radiation dose to mother and fetus which, depending on how the technique is performed, can vary by up to 40-fold. Of the techniques using X-rays, CT pelvimetry with a lateral scanogram generally gives the lowest radiation dose and conventional radiography using an air gap technique with a single lateral view is a relatively low-dose alternative where CT is not available. A questionnaire was sent to 227 hospitals during 1993 and 1996 to assess whether there was a move towards lower dose techniques of pelvimetry. The results show a trend away from conventional pelvimtry (48.4% in 1993 to 28% in 1996) with a small proportion of centres using MRI (4%) in 1996. Of the centres still using conventional pelvimetry, relatively few were using a low-dose air-gap technique (2.1% in 1993 to 10.9% in 1996). An increasing majority of centres were using one-view CT (69.3% in 1993 and 80.4% in 1996) but a significant proportion were still performing more than one view. This study shows that there was a move towards lower dose techniques of pelvimetry but that there were still many hospitals that had not implemented a policy of reducing radiation exposure in these patients.