Abstract
ConclusionsAn indirect calorimeter in which measurement of Vo2 is based on internal constant flow rather than spirometry can be used to accurately measure Vo2 from a continuous-flow breathing circuit, if the total circuit flow is less than the internal flow. This limitation may restrict the use of continuous flow to a level below the subject's peak inspiratory flow. The accuracy of indirect calorimetry cannot be guaranteed for any amount of tracheal tube cuff leak. (Crit Care Med 1992; 20:1335–1340)