A Novel Continuous Breath Test Versus Scintigraphy for Gastric Emptying Rate Measurement


    loading  Checking for direct PDF access through Ovid

Abstract

BackgroundGastric scintigraphy (GS) is considered the gold standard for gastric emptying rate (GER) measurement; however, it requires expensive equipment and special licensing for radioactive substances.AimTo compare a new nonradioactive, real-time, continuous breath test (CBT) method with GS for the GER measurement.MethodsSimultaneous GER analysis by both GS and CBT was carried out on 8 dyspeptic patients and 6 healthy controls. After a 14 hour fast, participants ate a standard meal of 250 Kcal double labeled with 1 mCi of Tc-99m and 100 μg of C-13 labeled octanoic acid. The participants underwent simultaneous GS for 120 minutes on single-detector γ camera and CBT carried out by attaching the nasal cannula of the system (Oridion, BreathID, Israel), which automatically and continuously collected and analyzed breath samples with real time display. A linear fit model was used to calculate gastric empting half-time. A half-time of more than 100 minutes for GS and 80 minutes for CBT were considered pathologic. The GS and CBT were compared by κ test of agreement in normal/abnormal results.ResultsGood correlation was found for GER measurements between GS and the CBT methods with a linear correlation coefficient of R=0.74. The κ test indicated excellent agreement with value of 0.86 for the qualitative determination of pathologic and normal results.ConclusionsThe novel CBT provides reliable and reasonably accurate data for on-line GER estimate, in a simple manner suitable for medical clinics or bedside setting, without the use of radioactive substances.