Twenty-four Hours of Sleep, Sedentary Behavior, and Physical Activity with Nine Wearable Devices

    loading  Checking for direct PDF access through Ovid

Abstract

Purpose

This study compared nine devices for accuracy in a 24-h activity measurement.

Methods

Adults (n = 40, 47% male) wore nine devices for 24 h: ActiGraph GT3X+, activPAL, Fitbit One, GENEactiv, Jawbone Up, LUMOback, Nike Fuelband, Omron pedometer, and Z-Machine. Comparisons (with standards) were made for total sleep time (Z-machine), time spent in SED (activPAL), LPA (GT3X+), MVPA (GT3X+), and steps (Omron). Analysis included mean absolute percent error, equivalence testing, and Bland–Altman plots.

Results

Error rates ranged from 8.1% to 16.9% for sleep, 9.5% to 65.8% for SED, 19.7% to 28.0% for LPA, 51.8% to 92% for MVPA, and 14.1% to 29.9% for steps. Equivalence testing indicated that only two comparisons were significantly equivalent to standards: the LUMOback for SED and the GT3X+ for sleep. Bland–Altman plots indicated GT3X+ had the closest measurement for sleep, LUMOback for SED, GENEactiv for LPA, Fitbit for MVPA, and GT3X+ for steps.

Conclusions

Currently, no device accurately captures activity data across the entire 24-h day, but the future of activity measurement should aim for accurate 24-h measurement as a goal. Researchers should continue to select measurement devices on the basis of their primary outcomes of interest.

Related Topics