Effect of variation in relaxation parameter value on LOR-RAMLA reconstruction of 18F-FDG PET studies


    loading  Checking for direct PDF access through Ovid

Abstract

PurposeWe tested the impact of different values of the relaxation parameter lambda (λ) on contrast and noise in line-of-response row-action maximum likelihood algorithm (LOR-RAMLA) in 18F-fluorodeoxyglucose positron emission tomography (PET)/computed tomography (CT) imaging.MethodsPhantom studies were performed on a Gemini XL PET/CT scanner. The NEMA/IEC (National Electrical Manufacturers Association/International Electro technical Commission) torso phantom was used and acquisition data were reconstructed with λ values ranging from 0.025 to 0.1. Quality of the reconstructed images was evaluated by contrast recovery coefficients and background variability values according to the NEMA NU 2-2001 procedures.ResultsContrast recovery coefficients and background variability increased significantly when λ increased. The best noise-versus-resolution trade-off was obtained with λ in the 0.04–0.06 range. For LOR-RAMLA reconstruction, the manufacturer allows a possible λ choice from 0.025 to 0.1. We would not advise too small (0.025) or too large (0.1) λ values which result in too smooth or too noisy images.ConclusionWe determined optimal λ values in LOR-RAMLA on a Gemini XL PET/CT scanner. Caution is needed when using λ values out of that range.