Comparison of continuous breath test and gastric scintigraphy for the measurement of gastric emptying rate in healthy and dyspeptic individuals


    loading  Checking for direct PDF access through Ovid

Abstract

ObjectiveTo determine normative gastric emptying rates for the continuous breath test and to compare its findings with gastric scintigraphy in healthy volunteers and dyspeptic patients.MethodsA standard 250 kcal meal double-labeled with 1 mCi 99mTc colloid and 100 μg nonradioactive 13C-octanoic acid was administered to 20 healthy individuals and 22 dyspeptic patients attending a tertiary medical center in 2009–2010. Gastric emptying rate was measured simultaneously with sequential gastric scintigraphy and the continuous breath test. The results of the healthy controls were used as a normative reference. The findings of the two tests were analyzed by linear regression and κ statistics. In addition, the gastric half-emptying times (T½) were compared by a κ test for evaluating the agreement of normal/abnormal results in both methods.ResultsBackground features were as follows: healthy individuals – 15 men/5 women, mean age 44.9±14 years and mean BMI 26.8±3.5; dyspeptic patients – 5 men/17 women, mean age 58.0±13 years and mean BMI 25.3±5.5. The upper limit of normal for gastric half-emptying time (T½) was 140 min by the breath test. The linear correlation of T½ between the methods was 0.64. The agreement of normal/abnormal T½ between the methods was 0.65.ConclusionThe continuous breath test may be a suitable method for the evaluation of gastric emptying. A larger study is required to determine its feasibility as the standard of care in dyspeptic patients.