Impact of applying the more stringent validation criteria of the revised European Society of Hypertension International Protocol 2010 on earlier validation studies


    loading  Checking for direct PDF access through Ovid

Abstract

ObjectiveSince 2002 when the European Society of Hypertension International Protocol (ESH-IP) was published it has become the preferred protocol for validating blood pressure monitors worldwide. In 2010, a revised version of the ESH-IP with more stringent criteria was published. This study assesses the impact of applying the revised ESH-IP criteria.MethodsA systematic literature review of ESH-IP studies reported between 2002 and 2010 was conducted. The impact of applying the ESH-IP 2010 criteria retrospectively on the data reported in these studies was investigated. The performance of the oscillometric devices in the last decade was also investigated on the basis of the ESH-IP criteria.ResultsAmong 119 published studies, 112 with sufficient data were analyzed. According to ESH-IP 2002, the test device failed in 19 studies, whereas by applying the ESH-IP 2010 criteria in 28 additional studies increased the failure rate from 17 to 42%. Of these 28 studies, in 20 (71%) the test device failed at part 1 (accuracy per measurement) and in 22 (79%) at part 2 (accuracy per subject). Most of the failures involved the ‘5 mmHg or less’ criterion. In the last decade there has been a consistent trend toward improved performance of oscillometric devices assessed on the basis of the ESH-IP criteria.ConclusionThis retrospective analysis shows that the stricter revised ESH-IP 2010 criteria will noticeably increase the failure rate of devices being validated. Oscillometric devices are becoming more accurate, and the revised ESH-IP by acknowledging this trend will allow more accurate devices to enter the market.

    loading  Loading Related Articles