Implantation of the Corinthian IQ stent into the femoropopliteal arteries using 6-F introducer sheaths in antegrade and crossover procedures: midterm results


    loading  Checking for direct PDF access through Ovid

Abstract

Abstract.The aim of this study was to evaluate the technical performance of the implantation of the Corinthian IQ stent into the femoropopliteal arteries using 6-F introducer sheaths. Thirty-three lesions (24 stenoses, 9 occlusions) in 32 patients (mean age 67 years) were stented. Selective stent implantation was performed after unsatisfactory percutaneous transluminal angioplasty result with an antegrade access in 21 patients and through a crossover sheath in 11 patients. A total of 36 stents were implanted. Stent delivery was successful in all cases (stent diameter 5 or 6 mm, stent length 11-38 mm). In none of the cases was stent dislodgement from the balloon observed. Stent placement was optimal without residual stenosis in 30 cases. In 1 case the stent was chosen too short requiring implantation of a second stent. In 1 patient distal embolization was observed which was treated with catheter embolectomy. Nine-month primary and secondary patency rates were 79±8.7 and 86±7.7%, respectively. Implantation of the Corinthian IQ stent into the femoropopliteal arteries using 6-F introducer sheaths is safe in antegrade and crossover procedures.