Profile of St. Jude Medical’s Allure Quadra quadripolar pacemaker system for cardiac resynchronization therapy

    loading  Checking for direct PDF access through Ovid

Abstract

Congestive heart failure is a major public health epidemic and economic burden in the USA and worldwide. Cardiac resynchronization therapy is an effective therapy for treating congestive heart failure in conjunction with pharmacologic therapy. The average congestive heart failure admission costs approximately US$ 8 billion annually. Current cardiac resynchronization therapy pacemaker systems from various manufacturers deliver therapy-utilizing bipolar leads including the left ventricle, with electrode spacing ranging from 8 to 22 mm. The Quartet LV™ lead model 1458Q (St. Jude Medical Sylmar, CA) is a quadripolar lead with a 4.0 Fr. tip electrode and three 4.7 Fr. ring electrodes located 20, 30 and 47 mm from the tip. The Quartet lead and Allure Quadra TM allows 14 pacing configurations, providing benefits, including reductions in phrenic nerve stimulation, reduced pacing thresholds, improved battery longevity and potential reductions, in non-responders to cardiac resynchronization therapy. In addition, there is cost benefit data from utilizing quadripolar technology compared with traditional bipolar cardiac resynchronization therapy.