Reproducibility of Optic Nerve Head Topography Measurements in Eyes with Undilated Pupils


    loading  Checking for direct PDF access through Ovid

Abstract

SUMMARYA confocal scanning diode laser ophthalmoscope (Retina Tomograph, Heidelberg Engineering, GmbH, Heidelberg, F.R.G.) was used to examine the optic nerve head in both normal healthy (n = 10) and glaucomatous (n = 10) eyes. An image series was obtained from 32 transverse optical sections taken at consecutive height planes (50–80 μm each) over a 1.5–2.5-mm scan depth. Each image was then analyzed to create a topographic map containing 256 ± 256 pixels (65,536 pixels) with afield of view of 10±. Images were obtained through undilated pupils. Reproducibility was 30.1 ± 7.0 and 31.8 ± 10.6 μm in normal and glaucomatous eyes, respectively. The instrument provides highly reproducible topographic data of the optic nerve head in patients with undilated pupils.