PWE-036 Use of an Ultraslim Paediatric Colonoscope to Complete Difficult Colonoscopic Procedures

    loading  Checking for direct PDF access through Ovid

Abstract

Introduction

Discomfort and failure to progress beyond the sigmoid are the commonest reasons for non-completion of colonoscopy.1 We anecdotally observed that the use of a new ultraslim paediatric colonoscope (Olympus PCF-PQ260L) when the standard colonoscope failed often led to completion of difficult cases.

Methods

We analysed 19 reports from cases at the Gloucestershire Royal hospital where the PCF-PQ260L was used as a second endoscope on an examination started with a regular colonoscope (Olympus H260 or Q260 colonoscopes) over the period January to August 2013.

Results

95% (18/19) of the time the extent of a non-completed colonoscopy was the sigmoid colon. 50% of the time this was due to diverticular disease with patient discomfort the second most common cause. In 80% of these cases, the subsequent use of the PCF-PQ260L enabled the endoscopist to reach the caecum. In addition, in those procedures that failed due to discomfort, comfort scores were improved in 50% during the second procedure with the PCF-PQ260L. The CIR of the GI consultants working at GRH averaged 96% for 2013.

Conclusion

The PCF-PQ260L enabled the negotiation of the sigmoid colon in 80% of cases where a standard endoscope failed in the hands of skilled colonoscopists. This limited study suggests that the PCF-PQ260L is an exciting new tool in the colonoscopist’s inventory.

Reference

1 Dafnis G, Granath F, Påhlman L, Ekbom A, Blomqvist P. Patient factors influencing the completion rate in colonoscopy. Dig Liver Dis. 2005 ;37(2):113–8

Disclosure of Interest

None Declared.