Comparison of range of motion after standard and high-flexion posterior stabilised total knee replacement

    loading  Checking for direct PDF access through Ovid

Abstract

The purpose of this study was to compare the range of motion after standard version posterior stabilised TKR and high-flexion version TKR in patients receiving bilateral total knee replacement. Thirty-five patients were recruited. The range of motion of the knees was measured clinically with a goniometer in both the pre-operative period and the most recent follow-up. It was found that the pre-operative range of motion was comparable in the two groups. The average post-operative flexion was 105° ± 13° in the standard version group and 106° ± 14° in the high-flexion design group (p = 0.201, paired t-test; beta error = 0.073). A slight loss in flexion was observed in the standard version group (0.5°) as opposed to a slight gain in the high-flexion design group (2°). However, this was not statistically significant (p = 0.251, paired t-test; beta error = 0.105).

Le but de cette étude est de comparer la mobilité du genou après prothèse totale postéro stabilisée et prothèse total type « high-flexion », tous les patients étant opérés des deux genoux dans le même temps opératoire. 35 patients ont été inclus, la mobilité des genoux a été mesurée cliniquement en pré-opératoire et au plus grand recul. La mobilité pré-opératoire a été comparable dans les deux groupes, la mobilité post opératoire a été de 105° ± 13° dans la prothèse standard contre 106° ± 14° dans les prothèses dites « high-flexion » (p = 0.201). Une petite perte de flexion est observée dans le groupe standard (0.5°), le gain dans le groupe des prothèses dites « high-flexion ’ est modéré (2°). Il n'y a pas de différences significatives entre ces deux groupes (p = 0.251).

Related Topics

    loading  Loading Related Articles