Percutaneous Mechanical Thrombolysis: A Comparative Study of Various Rotational Catheter Systems


    loading  Checking for direct PDF access through Ovid

Abstract

Schmitz-Rode T, Günther RW. Percutaneous mechanical thrombolysis: a comparative study of various rotational catheter systems. Invest Radiol 1991;26:557-563.In vitro fragmentation of thrombi was performed with the Kensey catheter, the Bildsoe catheter, the Günther aspiration catheter, and two modified aspiration catheters in a model circuit under femoral and pulmonary artery flow conditions. In the femoral artery model all tested catheters worked sufficiently: particles larger than 1000 µm in the effluent comprised only 1.27% ± 0.46% of the initial thrombus weight. In the treatment of larger thrombi (pulmonary artery model), the action of rotational aspiration devices, and in particular of the Bildsoe catheter, was limited: fibrin filaments winding around the rotating tip led to catheter obstruction. The Kensey catheter yielded best results in the fragmentation of large thrombi.