Computed Tomography Coronary Stent Imaging With Iterative Reconstruction: A Trade-off Study Between Medium Kernel and Sharp Kernel

    loading  Checking for direct PDF access through Ovid

Abstract

Objective

To evaluate the improvement of iterative reconstruction in image space (IRIS) technique in computed tomographic (CT) coronary stent imaging with sharp kernel, and to make a trade-off analysis.

Materials and Methods

Fifty-six patients with 105 stents were examined by 128-slice dual-source CT coronary angiography (CTCA). Images were reconstructed using standard filtered back projection (FBP) and IRIS with both medium kernel and sharp kernel applied. Image noise and the stent diameter were investigated. Image noise was measured both in background vessel and in-stent lumen as objective image evaluation. Image noise score and stent score were performed as subjective image evaluation.

Results

The CTCA images reconstructed with IRIS were associated with significant noise reduction compared to that of CTCA images reconstructed using FBP technique in both of background vessel and in-stent lumen (the background noise decreased by approximately 25.4% ± 8.2% in medium kernel (P Conclusions

Iterative reconstruction in image space reconstruction can effectively reduce the image noise and improve image quality. The sharp kernel images constructed with iterative reconstruction are considered the optimal images to observe coronary stents in this study.