Computed Tomography–Guided Tumor Ablation: Analysis and Optimization of Computed Tomography Technique With Various Ablation Devices

    loading  Checking for direct PDF access through Ovid

Abstract

Purpose

Ablation device-associated computed tomography beam hardening artifacts can make tumor margin visualization and its relationship to the ablation applicator tip challenging. Determine optimal scanning conditions for currently-used applicators.

Materials and Methods

Eleven applicators were placed in ex vivo cow livers with implanted mock tumors, surrounded by bolus gel. Various computed tomography scans were performed at 440 mA with 5 mm thickness changing kVp (80, 100, 120, 140), scan time (0.5, 0.7, 1.0, 2.0 seconds), adaptive statistical iterative reconstruction (ASiR) (30, 60, 90), scan type (helical, axial), pitch (0.5, 0.94, 1.37, 1.75), and reconstruction algorithm (soft, standard, lung). Two radiologists blindly scored the images for image quality and artifact quantitatively.

Results

Cool-tip single (CTS) RF electrode (Covidien) performed significantly better than all other devices in both perceived image quality and artifact while Boston Scientific 4.0 RF electrode (Boston Scientific) underperformed (all P < 0.001), when not controlling for any other factors. An effect for artifact (P < 0.001) was found for kVp and device: for most conditions, 80 kVp was rated significantly lower than all other levels, whereas 120 and 140 performed significantly better than 100 and 80. No significant effect with ASiR level and device was found for the artifact. There was an effect observed for artifact (P < 0.001) between scan time and probe: for most devices, 0.5 seconds was rated significantly lower than all other scan times, but CTS was resilient—showing no difference from other scan times. Algorithm did not show any significant effects. Taking into account ASiR, kVp, and time, CTS outperformed all other devices.

Conclusions

Higher kVp and scan times reduce device artifacts. It appears that CTS performs the best, even when considering ASiR, kVp, pitch, scan type, and scan time.