Accuracy of Automated Liver Contouring, Fat Fraction, and R2* Measurement on Gradient Multiecho Magnetic Resonance Images

    loading  Checking for direct PDF access through Ovid

Abstract

Objective

This study aimed to evaluate the performance of an automated workflow of volumetric liver proton density fat fraction (PDFFvol) and R2* quantification with automated inline liver volume (LV) segmentation.

Methods

Dual-echo and multiecho Dixon magnetic resonance images were evaluated in 74 consecutive patients (group A, PDFF < 10%; B, PDFF ≥ 10%; C, R2* ≥ 100 s−1; D, post-hemihepatectomy). The values of PDFFvol and R2*vol measurements across the LV were generated on multiecho images in an automated fashion based on inline liver segmentation on dual-echo images. Similar measurements were performed manually.

Results

Using the inline algorithm, the mis-segmented LV was highest in group D (80%). There were no significant differences between automated and manual measurements of PDFFvol. Automated R2*vol was significantly lower than manual R2*vol in group A (P = 0.004).

Conclusions

Inline LV segmentation performed well in patients without and with hepatic steatosis. In cases with iron overload and post-hemihepatectomy, extrahepatic areas were erroneously included to a greater extent, with a tendency toward overestimation of PDFFvol.