Inadvertent transposition of defibrillator coil terminal pins causing inappropriate ICD therapies

    loading  Checking for direct PDF access through Ovid

Abstract

We report the case of a 65-year-old man with chronic atrial fibrillation (AF) and severe ischemic cardiomyopathy who underwent implantation of a prophylactic single-chamber implantable cardioverter–defibrillator (ICD). The patient experienced inappropriate ICD therapies due to oversensing of pectoral muscle myopotential secondary to reversal of the defibrillator coil terminal pins in the ICD header. Recognizing this possibility is important to avoid misinterpretation of spontaneous oversensing as hardware failure (e.g., lead fracture or insulation breech) and potentially unnecessary ICD system surgical intervention, including lead extraction.