Incidence of delivery wire recapture failure with the Pipeline Flex device

    loading  Checking for direct PDF access through Ovid

Abstract

Background

The Pipeline Flex embolization device has several advantages over the first-generation Pipeline embolization device (Covidien, Dublin, Republic of Ireland). Despite these advantages, we have observed frequent difficulty in recapturing the device's delivery wire.

Objective

To prospectively document the incidence of failure to recapture the delivery wire.

Methods

We tracked our experience in patients undergoing endovascular treatment with a flow-diverting stent for cerebral aneurysms between 1 May and 30 September 2016. Patient and lesion characteristics, device dimensions, and technical outcomes of delivery wire recapture were prospectively recorded for each device.

Results

Eighteen devices were deployed in 15 patients by the senior author (FCA) during this period. Failure to recapture the delivery wire occurred in 10 of 18 (56%) cases. No adverse outcomes of delivery wire recapture failure were encountered in this series.

Conclusions

The incidence of delivery wire recapture failure with the Pipeline Flex device is high. Failure to recapture the delivery wire carries a theoretical risk of stent displacement when re-navigating across the device, and endovascular surgeons should be aware of this limitation. Authors are encouraged to report delivery wire recapture failure rates in future clinical series in which the Pipeline Flex device is used.