Assessment of human placenta as an ex-vivo vascular model for testing of liquid embolic agent injections with adjunctive techniques

    loading  Checking for direct PDF access through Ovid

Abstract

Purpose

This project sought to test the utility of post-delivery human placenta (HP) as a vascular model for liquid embolic agent (LEA) simulation, along with adjunctive techniques.

Materials and methods

Twelve LEA injections were performed under fluoroscopy in HP with two reflux control methods: dual lumen ‘mini’ balloon-catheter (n=9); and injection after proximal nBCA plug formation through a second microcatheter (‘pressure cooker’) (n=3). Measured outcomes included liquid embolic agent (LEA) advancement and reflux. Reflux was categorized into three grades: grade 0=no reflux; grade 1=occlusion of side branches without reflux beyond the balloon or plug; and grade 2=reflux beyond the balloon or plug.

Results

Simulation success was greater when a balloon was used rather than with a nBCA plug (89% vs 33%, P=0.054). In eight successful balloon-assisted injections, the reflux grades were: 50% grade 0; 12.5% grade 1; and 37.5% grade 2. The one successful nBCA plug injection had grade 2 reflux. All grade 2 balloon injections occurred when the balloon was positioned across a vessel bifurcation.

Conclusions

HP provides excellent simulation for liquid embolic agents with a dual lumen balloon catheter.