Development and application of a novel, dual-mode gradient, stability-indicating HPLC-DAD method for the simultaneous determination and purity assessment of mebeverine hydrochloride, diloxanide furoate and their corresponding major degradation products in combination with some gastrointestinal drugs in the form of oral doses

    loading  Checking for direct PDF access through Ovid

Abstract

A simple, precise, rapid, stability-indicating reversed phase high performance liquid chromatographic method with photodiode array detection was developed and validated for the determination of mebeverine hydrochloride in combination with sulpiride or with diloxanide furoate and metronidazole in the presence of their corresponding degradation products. Optimum separation was achieved in less than 10 min using an X-Bridge C18 column (150 mm × 4.6 mm i.d., 3.5 μm particle size); elution was accomplished via the application of a dual-mode solvent and flow rate gradient system. This elution system enables the separation of nine components within a cycle time of 15 min and with a resolution greater than 2.5. Detection was conducted at 230 nm, and purity assessment was performed using a photodiode array detector. The method has been validated with respect to specificity, linearity, accuracy, precision, limit of quantitation, limit of detection, robustness and ruggedness. The validation criteria were met in all cases. The developed HPLC method was successfully applied to commercial tablets. It was shown that this method is very sensitive to the determination of the degradation products, downward to 0.1 w/w% levels, which is far below the limits for testing these degradation products within their corresponding intact drugs.

Graphical abstract

Diagram of dual-mode gradient.

Related Topics

    loading  Loading Related Articles