Automation of plasma protein binding assay using rapid equilibrium dialysis device and Tecan workstation

    loading  Checking for direct PDF access through Ovid

Abstract

Binding of drug molecules to plasma proteins is an important parameter in assessing drug ADME properties. Plasma protein binding (PPB) assays are routinely performed during drug discovery and development. A fully automated PPB assay was developed using rapid equilibrium dialysis (RED) device and Tecan workstation coupled to an automated incubator. The PPB assay was carried out in unsealed RED plates which allowed the assay to be fully automated. The plasma pH was maintained at 7.4 during the 6-h dialysis under 2% CO2 condition. The samples were extracted with acetonitrile and analyzed by liquid chromatography tandem mass spectrometry. The percent bound results of 10 commercial drugs in plasma protein binding were very similar between the automated and manual assays, and were comparable to literature values. The automated assay increases laboratory productivity and is applicable to high-throughput screening of drug protein binding in drug discovery.