Restricted Field of View Magnetic Resonance Imaging of a Dynamic Time Series

    loading  Checking for direct PDF access through Ovid

Abstract

A restricted field of view (rFOV) approach for imaging a dynamic time series of volumes of limited spatial extent within a larger subject is described. The shorter readout with rFOV-MRI can be exploited to either limit image artifacts or increase spatial resolution. To accomplish rFOV imaging of a multislice volume for a dynamic series, an outer volume suppression (OVS) preparation that saturates signal external to a cylinder through the subject is followed by slice-selective excitation and a spiral readout. The pass- and stopband efficiencies of the OVS in an agar gel phantom were 97% (±1.5%) and 3% (±1%), respectively. Profiles of the temporal signal-to-noise ratio (SNR) were measured in a phantom and an adult brain. The rFOV sequence reduced distortions from off-resonance signal andTSymbol-induced blurring compared to a conventional sequence. Sequence utility is demonstrated for high-resolution rFOV functional MRI (fMRI) in the visual cortex. The rFOV sequence may prove to be useful for other multislice dynamic and high-resolution imaging applications. Magn Reson Med 57:297–307, 2007. © 2007 Wiley-Liss, Inc.