Optimizing the Precision-Per-Unit-Time of Quantitative MR Metrics: Examples forT1,T2, and DTI

    loading  Checking for direct PDF access through Ovid

Abstract

Quantitative MR metrics (e.g.,T1,T2, diffusion coefficients, and magnetization transfer ratios (MTRs etc)) are often derived from two images collected with one acquisition parameter changed between them (the “two-point” method). Since a low signal-to-noise-ratio (SNR) adversely affects the precision of these metrics, averaging is frequently used, although improvement accrues slowly—in proportion to the square root of imaging time. Fortunately, the relationship between the images' SNRs and the metric's precision can be exploited to our advantage. Using error propagation rules, we show that for a given sequence, specifying the total imaging time uniquely determines the optimal acquisition protocol. Specifically, instead of changing only one acquisition parameter and repeating the imaging pair until all available time is spent, we propose to adjustallof the parameters and the number of averages at each point according to their contribution to the sought metric's precision. The tactic is shown to increase the precision of the well-known two-pointT1,T2, and diffusion coefficients estimation by 13–90% for the same sample, sequence, hardware, and duration. It is also shown that under this general framework, precision accrues faster than the square root of time. Tables of optimal parameters are provided for various experimental scenarios. Magn Reson Med 57:380–387, 2007. © 2007 Wiley-Liss, Inc.