Reduction of scatter-induced image noise in cone beam computed tomography: effect of field of view size and position

    loading  Checking for direct PDF access through Ovid

Abstract

Objective

To measure the effect of field of view (FOV) size and position on scatter-induced image noise in cone beam computed tomography (CBCT).

Study Design

A polymethyl methacrylate (PMMA) phantom containing air and aluminum underwent CBCT scanning, using seven FOVs ranging between 4 × 4 cm and 14 × 10 cm, positioned both centrally and according to a dental scan. Signal difference to noise ratio (SDNR) was measured on two-dimensional (2-D) projection images.

Results

At a central position, SDNR decreased with increasing FOV size, ranging between 9.8 (14 × 10 cm) and 10.9 (4 × 4 cm). For dental FOV positions, SDNR values were between 6.3 (14 × 10 cm) and 9.5 (4 × 4 cm). To reach a constant SDNR, a dose reduction up to 76% was possible for small FOVs compared with the 14 × 10 cm FOV.

Conclusions

The use of small FOVs and peripheral FOV positioning decreases scatter at the detector, resulting in a considerable potential for reduction of radiation dose to the patient.