Transvenous Dual Site Left Ventricular Pacing to Lower Elevated Pacing Thresholds


    loading  Checking for direct PDF access through Ovid

Abstract

Dual site left ventricular pacing through two left ventricular pacing leads, located in discrete vessels, significantly lowered pacing thresholds from 6 V at 1 ms and 4.25V at 0.5 ms through the leads individually, to 0.75V at 0.5 ms by utilizing a Y-adaptor to connect the two leads. (PACE 2011; e6-e8)