A randomized comparison of pediatric-sized Streamlined Liner of Pharyngeal Airway™ and Laryngeal Mask Airway-Unique™ in paralyzed children

    loading  Checking for direct PDF access through Ovid

Abstract

Background:

The pediatric-sized Streamlined Liner of Pharyngeal Airway™ (SLIPA) is a new supraglottic airway device for children.

Aims:

The aim of this study was to compare the clinical performance of the pediatric-sized SLIPA with the Laryngeal Mask Airway-Unique in paralyzed children under positive pressure-controlled ventilation (PCV).

Methods:

One hundred children, aged 2 months to 12 years with American Society of Anesthesiologists physical status I–II were enrolled and randomly allocated to the SLIPA group or the Laryngeal Mask Airway-Unique group (50 patients in each group). The primary outcome variable was oropharyngeal leak pressure. Other outcome variables were first insertion success rate, insertion time, minor airway interventions required for successful insertion, intraoperative dislodgement, ventilatory data, and perioperative complications.

Results:

The insertion characteristics, ventilation data, and perioperative complications were comparable between the two groups. The leak pressure of the SLIPA was significantly higher than that of the Laryngeal Mask Airway-Unique [median (IQR): 25 (22–30) cmH2O vs 21 (19–26) cmH2O, respectively; mean ± sd: 25.3 ± 4.6 cmH2O vs 22.6 ± 4.8 cmH2O, respectively; P = 0.006]. The incidence of intraoperative dislodgment was significantly lower in the SLIPA group than in the Laryngeal Mask Airway-Unique group (0 vs 6 patients, respectively; P = 0.027).

Conclusions:

In conclusion, both the SLIPA and the Laryngeal Mask Airway-Unique can be used effectively without severe complications in paralyzed children. Additionally, the SLIPA provides a better airway seal and better intraoperative position stability than the Laryngeal Mask Airway-Unique.