High resolution CBV assessment with PEAK-EPI: k-t-undersampling and reconstruction in echo planar imaging

    loading  Checking for direct PDF access through Ovid

Abstract

Purpose:

Achieving higher spatial resolution and improved brain coverage while mitigating in-plane susceptibility artifacts in the assessment of perfusion parameters, such as cerebral blood volume, in echo planar imaging (EPI)-based dynamic susceptibility contrast weighted cerebral perfusion measurements.

Methods:

PEAK-EPI, an EPI sequence with interleaved readout trajectories and three different strategies for autocalibration-signal acquisition (inplace, dynamic extra and extra) is presented. Performance of each approach is analyzed in vivo based on flip angle variation induced dynamics, assessing temporal fidelity, temporal SNR and g-factors. All approaches are compared with conventional GRAPPA reconstructions. PEAK-EPI with inplace autocalibration-signal at R = 5 is then compared with the standard clinical EPI protocol in six patients, using two half-dose dynamic susceptibility contrast weighted cerebral perfusion measurements per subject.

Results:

PEAK-EPI acquisition facilitates a substantial increase of spatial resolution at a higher number of slices per TR and provides improved SNR compared to conventional GRAPPA. High dependency of the resulting reconstruction quality on the type of autocalibration-signal acquisition is observed. PEAK-EPI with inplace autocalibration-signal achieves high temporal fidelity and initial feasibility is shown.

Conclusion:

The obtained high resolution cerebral blood volume maps reveal more detailed information than in corresponding standard EPI measurements and facilitate detailed delineation of tumorous tissue. Magn Reson Med, 2016. © 2016 Wiley Periodicals, Inc.