MRI audit of complications in intracranial stenosis treated with Wingspan device

    loading  Checking for direct PDF access through Ovid

Abstract

Objectives

To evaluate the safety and efficacy of the Wingspan device for the treatment of symptomatic intracranial atherosclerotic stenosis (ICAS).

Methods

We audited a prospective ongoing database of consecutive patients who received Wingspan stenting between January 2013 and December 2015. All patients underwent MRI to audit any complications during the early follow-up period. We focused on the clinical demographics, lesion characteristics, treatment results, and periprocedural complications. Functional outcomes were measured with the modified Rankin Scale (mRS) at discharge and after 3 months.

Results

Intracranial stenting was performed in 50 patients (100%). Mean stenosis pre-stenting was 76.5±13.1% and post-stenting residual stenosis was 19.8±13.8%. The overall 30-day rate of procedure-related complications was 6.0% (3/50). Two patients (4%) developed in-stent restenosis, one of whom had a dissection at the middle cerebral artery. Interestingly, on the follow-up MRI scan there was a high incidence of asymptomatic diffusion-weighted imaging (DWI) hyperintensities, 46% (23/50) presumed to be due to microembolic causes. At the 90-day, 180-day, and 1-year follow-up, three patients had further strokes resulting in a total complication rate of 12%. 92% had excellent outcomes (mRS 0–1) and only one patient had deterioration of his mRS score.

Conclusions

ICAS treated by Wingspan stenting using pre-placement balloon angioplasty appears safe and effective with a high technical success rate and favorable outcomes. There is a high incidence of asymptomatic DWI hyperintensites post-procedure, but these do not appear to result in long-term sequelae.

Related Topics