A UHPLC method for the rapid separation and quantification of anthocyanins in acai berry and dry blueberry extracts

    loading  Checking for direct PDF access through Ovid

Abstract

The presented work describes the development and validation of a rapid UHPLC-UV method using a core-shell particle column with a pentafluorophenyl stationary phase for the separation and quantitative analysis of the six anthocyanins in acai berry and dry blueberry extracts. The anthocyanins (cyanidin-3-glucoside, cyanidin-3-rutenoside, delphinidin-3-galactoside, delphinidin-3-glucoside, delphinidin-3-rutenoside, and peonidin-3-glucoside) were separated and analyzed in 5 min. The chromatographic separation was performed on a Kinetex PFP (150 × 2.1 mm) core-shell column with a particle size of 1.7 μm at a temperature of 50 °C. Acetonitrile was used as mobile phase B and 5% formic acid, filtrated through a 0.22 μm filter, as mobile phase A. They were delivered at a flow rate of 0.55 mL min−1 according to the elution gradient program. The detection wavelength was set at 520 nm. A solid-liquid extraction with a solution of methanol and a 5% water solution of formic acid (25 + 75 v/v) using an ultrasonic bath was chosen for the preparation of the available commercial samples of food supplements with a content of acai berry extract and blueberry extract. Under optimal chromatographic conditions, the method was validated. Recoveries for all analyzed anthocyanins were 97.8–102.6% and the relative standard deviation ranged from 0.4% to 3.0% for within-day and from 0.6% to 3.1% for between-day repeatability. The limits of detection were in the range of 0.11–0.14 μg mL−1.