Validation of the YuWell YE690A upper-arm blood pressure monitor, for clinic use and self-measurement, according to the European Society of Hypertension International Protocol revision 2010

    loading  Checking for direct PDF access through Ovid

Abstract

Objective

The present study aimed to evaluate the accuracy of the automated oscillometric upper-arm blood pressure monitor YuWell YE690A for blood pressure measurement according to the International Protocol of the European Society of Hypertension revision 2010.

Participants and methods

Systolic and diastolic blood pressures were measured sequentially in 33 adult Chinese (12 women, 44.2 years of mean age) using a mercury sphygmomanometer (two observers) and the YE690A device (one supervisor). A total of 99 pairs of comparisons were obtained from 33 participants for judgments in two parts with three grading phases.

Results

All the blood pressure requirements were fulfilled. The YuWell YE690A device achieved the targets in part 1 of the validation study. The number of absolute differences between device and observers within 5, 10, and 15 mmHg was 79/99, 96/99, and 97/99, respectively, for systolic blood pressure and 72/99, 95/99, and 98/99, respectively, for diastolic blood pressure. The device also fulfilled the criteria in part 2 of the validation study. Thirty-one and 25 participants for systolic and diastolic blood pressure, respectively, had at least two of the three device–observer differences within 5 mmHg (required ≥24). No participant for systolic and two participants for diastolic blood pressure had all the three device–observer comparisons greater than 5 mmHg.

Conclusion

The YuWell blood pressure monitor YE690A has passed the requirements of the International Protocol revision 2010 and hence can be recommended for blood pressure measurement in adults.