Performance of Robotic Assistance for Skull Base Biopsy: A Phantom Study

    loading  Checking for direct PDF access through Ovid

Abstract

Objectives

This study aims to evaluate the feasibility of a custom robot system guided by optical cone beam computed tomography (CBCT)-based navigation for skull base biopsy.

Design

An accuracy study was conducted.

Setting

Platform for navigation and robot-aided surgery technology.

Participants

Phantom skull.

Main Outcome Measures

The primary outcome measure was to investigate the accuracy of robot-assisted needle biopsy for skull base tumors. A 14-gauge needle was automatically inserted by the five degrees of freedom robot into the intended target, guided by optical navigation. The result was displayed on the graphical user interface after matrix transformation. Postoperative image scanning was performed, and the result was verified with image fusion.

Results

All 20 interventions were successfully performed. The mean deviation of the needle tip was 0.56 ± 0.22 mm (measured by the navigation system) versus 1.73 ± 0.60 mm (measured by image fusion) (p < 0.05). The mean insertion depth was 52.3 mm (range: 49.7-55.2 mm). The mean angular deviations off the x-axis, y-axis, and z-axis were 1.51 ± 0.67, 2.33 ± 1.65, and 1.47 ± 1.16 degrees, respectively.

Conclusions

The experimental results show the robot system is efficient, reliable, and safe. The navigation accuracy is a significant factor in robotic procedures.