Validation of the Pangao PG-800A36 automatic wrist blood pressure monitor according to the European Society of Hypertension and the British Hypertension Society protocols

    loading  Checking for direct PDF access through Ovid

Abstract

Objective

The aim of this study was to validate the PG-800A36 automatic wrist blood pressure monitor according to the European Society of Hypertension International Protocol (ESH-IP) revision 2010 and the British Hypertension Society (BHS) protocols.

Participants and methods

A total of 33 participants were initially included on the basis of the ESH-IP, followed by examination of 85 participants according to the BHS protocol. The procedures and analysis methods of the protocols were followed precisely with left arm/wrist sequential measurements by two trained observers using a mercury sphygmomanometer and one supervisor using the device.

Results

The device passed the ESH-IP with an average difference of 1.45±6.46 mmHg for systolic blood pressure and 1.25±5.10 mmHg for diastolic blood pressure. Furthermore, the A/A grade of the BHS protocol was achieved with an average difference of 1.84±6.94 mmHg for systolic blood pressure and 1.15±6.49 mmHg for diastolic blood pressure, and thus, the device also fulfilled the requirements of the Association for the Advancement of Medical Instrumentation.

Conclusion

The Pangao PG-800A36 passed the requirements of the ESH-IP revision 2010 and achieved the A/A grade of the BHS protocol, which can be recommended for self-measurement in the general population.