Intra- and Interexaminer Variability of Two Separate Video Head Impulse Test Systems Assessing All Six Semicircular Canals

    loading  Checking for direct PDF access through Ovid

Abstract

Objective:

To evaluate intra- and interexaminer variability of the video Head Impulse Test (v-HIT) when assessing all six semicircular canals (SCCs) of two separate v-HIT systems.

Study Design:

Prospective study.

Setting:

Department of Otolaryngology, Head and Neck Surgery, Aalborg University Hospital, Denmark.

Patients:

One hundred twenty healthy subjects.

Intervention:

Four separate tests of all six SCCs with either system A or system B. Two examiners tested all subjects twice. Pretest randomization included type of v-HIT system, order of paired SCC testing, as well as initial examiner.

Intervention:

Main Outcome Measure: Gain values and the presence of pathological saccades were registered. Ninety-five percent limits of agreement (LOAs) were calculated for both intra- and interexaminer variability. Adding or subtracting the value from the mean difference achieves the upper and lower bound LOA. Ninety-five percent of the differences lie within these limits.

Results:

Interexaminer reliability: System A: LOAs between 0.13 and 0.24 for the horizontal SCCs and between 0.42 and 0.74 for the vertical SCCs. System B: LOAs between 0.09 and 0.13 for the horizontal SCCs and between 0.13 and 0.20 for the vertical SCCs. Intraexaminer reliability: System A: LOAs were 0.19 and 0.14 for the horizontal SCCs and varied from 0.43 to 0.53 for the vertical SCCs. System B: LOAs were 0.14 for the horizontal SCCs and varied from 0.13 to 0.22 for the vertical SCCs.

Conclusion:

Horizontal SCC testing: both v-HIT systems displayed good intra- and interexaminer variability. Vertical SCC testing: System B displayed good intra- and interexaminer variability whereas the opposite was true with system A.