Image Quality, Overall Evaluability, and Effective Radiation Dose of Coronary Computed Tomography Angiography With Prospective Electrocardiographic Triggering Plus Intracycle Motion Correction Algorithm in Patients With a Heart Rate Over 65 Beats Per Minute

    loading  Checking for direct PDF access through Ovid

Abstract

Purpose:

Recently, a new intracycle motion correction algorithm (MCA) was introduced to reduce motion artifacts from heart rate (HR) in coronary computed tomography angiography (cCTA). The aim of the study was to evaluate the image quality, overall evaluability, and effective radiation dose (ED) of cCTA with prospective electrocardiographic (ECG) triggering plus MCA as compared with standard protocol with retrospective ECG triggering in patients with HR≥65 bpm.

Materials and Methods:

One hundred consecutive patients (67±10 y) scheduled for cCTA with 65Results:

cCTA was successfully performed in all patients. In a segment-based model, group 1 compared with group 2 showed a lower rate of overall artifacts (67% vs. 83%; P<0.001) and motion artifacts (49% vs. 66%; P<0.001), resulting in a better Likert image quality score (2.83±1.03 vs. 2.37±1.02; P<0.01) and overall evaluability (85% vs. 75%; P<0.01). Group 1 showed a lower ED as compared with group 2 (3.1±1.9 vs. 11.9±3.3 mSv; P<0.01).

Conclusion:

MCA and cCTA with prospective ECG-triggering acquisition in patients with high HR improves image quality and overall evaluability compared with cCTA with standard retrospective ECG triggering.