Intraoperative Measurement for a New Transcutaneous Bone Conduction Hearing Implant


    loading  Checking for direct PDF access through Ovid

Abstract

ObjectiveTo investigate the possibility of using a modified reverse transfer function (RTF) measurement intraoperatively during surgery of a new transcutaneous bone conduction hearing implant to evaluate the status of the device.MethodsTests were performed on a cadaver skull (preclinically) and two conductive hearing loss patients implanted with a new transcutaneous bone conduction implant. During intraoperative activation, the RTF was measured using a microphone attached perpendicularly and directly to the skin in the middle section of the forehead.ResultsThe RTF could be measured for all frequencies from 500 to 6, 000 Hz.ConclusionThe usage of an intraoperative RTF measurement may be a good method to verify the mechanical coupling of the bone conduction floating mass transducer and to test the functional integrity of the implant in an objective way.