OBJECTIVE TASK-BASED ASSESSMENT OF LOW-CONTRAST DETECTABILITY IN ITERATIVE RECONSTRUCTION

    loading  Checking for direct PDF access through Ovid

Abstract

Evaluating image quality by using receiver operating characteristic studies is time consuming and difficult to implement. This work assesses a new iterative algorithm using a channelised Hotelling observer (CHO). For this purpose, an anthropomorphic abdomen phantom with spheres of various sizes and contrasts was scanned at 3 volume computed tomography dose index (CTDIvol) levels on a GE Revolution CT. Images were reconstructed using the iterative reconstruction method adaptive statistical iterative reconstruction-V (ASIR-V) at ASIR-V 0, 50 and 70 % and assessed by applying a CHO with dense difference of Gaussian and internal noise. Both CHO and human observers (HO) were compared based on a four-alternative forced-choice experiment, using the percentage correct as a figure of merit. The results showed accordance between CHO and HO. Moreover, an improvement in the low-contrast detection was observed when switching from ASIR-V 0 to 50 %. The results underpin the finding that ASIR-V allows dose reduction.