Evaluation of a Digital Flocculation Reader for the Rapid Plasma Reagin Test for the Serological Diagnosis of Syphilis

    loading  Checking for direct PDF access through Ovid

Abstract

We described the ASiManager-AT digital flocculation reader to demonstrate concordance between visual and digital readings of the rapid plasma reagin test for detection of antibodies in the serum of patients with syphilis. A qualitative and quantitative rapid plasma reagin was performed on each serum samples giving a concordance of 98.6% and 99.7%, respectively, for reactives and 100% for nonreactives.

Related Topics