Abbott AxSYM Vancomycin II Assay: Multicenter Evaluation and Interference Studies

    loading  Checking for direct PDF access through Ovid

Abstract

Abstract:

The authors evaluated the performance characteristics of the Abbott AxSYM Vancomycin II immunoassay in sera of patients with (n = 93 samples) and without (n = 327 patients) renal dysfunction. Correlation of vancomycin measurements with the Abbott AxSYM Vancomycin, Abbott TDx/TDxFLx, Syva enzyme-multiplied immunoassay technique (EMIT), DuPont automated chemistry analyzer (ACA), and high-performance liquid chromatography methods showed acceptable correlation as indicated by: slope values >0.95, r-values >0.97, y-intercepts <1.7 µg/ml, and Sy/x ranging from 9% to 15% of the average vancomycin value. The AxSYM Vancomycin II assay showed acceptable correlation with AxSYM vancomycin, TDx/TDxFLx, and high-performance liquid chromatography methods in 93 samples from patients with renal dysfunction. This monoclonal antibody-based assay showed no apparent interference from the presence of human antimouse antibody (HAMA) or the microbiologically inactive vancomycin crystalline degradation product (CDP). The authors conclude that the AxSYM Vancomycin II assay showed satisfactory agreement with other methods tested in this study.